Skip to main content

16. Warnings, Errors and Troubleshooting

If the instrument encounters a problem, it enters a different state, depending on the severity of the issue. There are warnings, noncritical errors and critical errors, each having different implications and requiring different actions.

16.1 Warnings​

Warnings appear when unusual behaviour is detected during an otherwise successful measurement. They can indicate reduced accuracy of the measurement results or indicate impending errors. Unlike errors, warnings do not prevent the instrument from functioning, but users should pay attention to them as they can indicate sources of inaccuracies.

WARNINGIMAGE
When unusual behavior is detected, a warning is shown next to the measurement results. More information on the warning can be found by clicking on the corresponding warning entry next to the measurement results or directly in the warning log.BactoSense — High SSC baseline warning

The following warning messages can be displayed:

Table 16-1 Warnings

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
W01Cartridge expiredCartridge reagents have expired. Measurement accuracy could be affected.• Please replace the cartridge as soon as possible.
W02-04High baseline signal in FL1 / FL2 / SSCBaseline signals are too high. This is usually due to bubbles or particles in suspension in the sample. Measurement accuracy could be affected.• Check the inlet tubing for any signs of deposits.
• Check the water connections, be sure that they are all tight (no air entry).
• If the warning persists, please perform cleaning of instrument with the cleaning kit.
W05Incubator temperature off targetTemperature during incubation was more than 2 °C off target. Cell counts may be affected. The incubator may be defective.• If the problem persists, please contact customer service.
W06No events detectedNo events detected during the measurement.• If the problem persists, please contact customer service.
W07TCC out of rangeTCC is above the specification limit of BactoSense of 2’000’000 cells/ml.• Please dilute the sample to ensure accuracy of the measurement.
W08Air bubbles detectedThe sample appears to contain air bubbles.• Check sampling device is installed properly.
• Check the sample for bubbles.
• If using online sampling device, check water is supplied at correct pressure.
W10Enclosure too coldEnclosure is too cold, rapid heating has been activated before continuing protocol.• Instrument cannot safely operate if internal temperature is too cold.
• Rapid heating attempts to heat the enclosure to ensure safe operations.
• Measurements will start once internal temperature is sufficient.
W11Service requiredService of the instrument is required. Measurement precision can no longer be guaranteed. Continued operation can lead to leaks, which can damage the instrument.• Next service date is overdue.
• Valve and plunger have reached end of life.
• Please contact your service representative.
W13Laser end of lifeLaser has reached end of life.• Please contact your service representative.
W14Non-critical power cutThe instrument was powered off unexpectedly.• Ensure the cleanliness of the instrument.
• If the issue occurs frequently, install an uninterruptible power supply

16.2 Non-critical errors​

Non-critical errors prevent a measurement from terminating successfully, but do not prevent the instrument from running another measurement afterwards. These errors do not require human intervention. Some non-critical errors are promoted to critical errors if they repeat three times.

NON-CRITICAL ERRORS
The protocol stops. The cause of the error is usually fixed by repeating the analysis or waiting.
• Another measurement can immediately be started. If it completes successfully, the error state is cleared.
• More information on the error can be found in the error log, by clicking on the corresponding error entry.
BactoSense — Non-critical empty sample error

The following non-critical error messages can be displayed:

Table 16-2 List of non-critical errors.

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
E01Cartridge door openCartridge door is open and prevents measurements from running.• Close door and retry the measurement.
E08Enclosure too dampEnclosure humidity is too high for safe operation.• Replace desiccant bag, by unscrewing the large cap on the left-hand-side of the instrument (see Replacement of the desiccant bag).
• If the problem persists, please contact customer service.
E09System overheatedInside temperature is too high for safe operation.• Reduce ambient temperature or increase measurement interval to allow instrument to cool down.
E14Laser too hotLaser temperature is too high for safe operation.• Reduce ambient temperature or increase measurement interval to allow instrument to cool down.
E15Mixer inflation errorUnderpressure detected during mix or dispense.• Missing sample.
• Air in the input.
• Leak in the system.
• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
E18Pump underpressure errorUnderpressure detected during dispense.• Missing sample.
• Air in the input.
• Leak in the system.
• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
E19System overheatedThe inner temperature is too high for operation.• Reduce ambient temperature, or increase measurement interval to allow instrument to cool down.
E33Processing memory overflowSignal processing circuits memory has overflown.• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
• Retry the measurement.
E42Enclosure too cold despite heatingEnclosure temperature remains too cold after heating procedure.• Ambient temperature is less than 5 °C (41 °F).
• Increase ambient temperature.
E44Pump module startup errorThe pump module failed to power on correctly.• Simply retry the measurement.
E45Low valid volume ratio errorThe valid volume ratio is too low. This can be caused by a too elevated bacterial content in the water or turbidity/debris.• The accuracy of the measurement is probably very low. Repeat the measurement while diluting the sample with clear water.
E46Cartridge is still busyThe cartridge was still initializing when the measurement started.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E48Microfluidic subsystem busyThe microfluidic controller is performing another task while the measurement was taken.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E49GPIO Timeout errorThe processing board is waiting for another module to respond.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E50Pump motion blockedThe motion of the pump is blocked.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E51Unable to get temperature readingUnable to get temperature reading. The control loop is thus disabled.• Non-critical error. Repeat the measurement. If the problem persists, contact a service technician.
E52Cartridge valve motion blockedThe motion of the cartridge valve is blocked.• Non-critical error. Repeat the measurement. If the problem persists, contact customer service.
E55M7 not connectedThe acquisition processor cannot be initialized.• Reboot the instrument.
• If the problem persists, contact customer service.
E56Empty acquisitionNo events detected during acquisition• Clean the sampling device using the Washstation, then perform a measurement of clean water.
• If the problem persists, contact customer service.
E57Cartridge busy (special command busy)Communication with the cartridge failed.• Non-critical error. Repeat the measurement.
• If the problem persists, contact customer service.
E58Cartridge busy (timeout)Communication with the cartridge failed.• Check the electronic connector between the cartridge and the instrument.
• Repeat the measurement.
• If the problem persists, contact customer service.
E59Valve sensor errorUnable to read valve position.• Please contact customer service.

16.3 Critical errors​

Critical errors interrupt the workflows of the instrument and require human intervention. The error source needs to be eliminated and the instrument set to running conditions again. Critical errors need to be cleared.

CRITICAL ERRORSIMAGE
If a critical error occurs during operation, it has the following effects:
• The protocol immediately stops.
• The instrument goes into a critical error state, and manual intervention is needed before any new protocol can be launched.
• The cause of the error must be solved by an operator. The errors can then be manually cleared from the Error log, as described in Error and warning logs and Clearing errors.
BactoSense — Critical missing cartridge error

The following critical error messages can be displayed:

Table 16-3 List of critical errors.

ERROR CODENAMEDESCRIPTIONCAUSES / WHAT TO DO
E00General errorUnhandled error.• Please contact customer service
E02Cartridge emptyCartridge is empty.• Replace cartridge (see Change cartridges)
E04Cartridge missingCartridge is missing.• Insert a cartridge (see Installing a cartridge)
E05Full waste bagCartridge waste bag is full.• Replace cartridge (see Change cartridges)
E06Cartridge communication impossibleCartridge is disconnected, door is open, or cartridge electronics are damaged.• Make sure the cartridge’s electronic cable is connected properly, and that the door is properly closed.
• If that doesn’t help, replace the cartridge and inform customer service.
E07Storage disk fullInsufficient storage space to continue operation.• Delete old measurements and try again (see Clearing of disk space)
• This will delete the measurements permanently from the device.
• Export measurements as backup solution.
E12FPGA initialization errorFailed to initialize the signal processing chip.• Reboot the instrument.
E16Processing errorAn error was discovered while processing the signals.• The accuracy of the latest result is not guaranteed. Please repeat the analysis.
E17Overpressure detectedOverpressure detected during dispense.• Likely cause: The output filter is probably clogged.
• Optical flow cell blocked.
• Please call customer service.
E20Abnormal shutdownProtocol interrupted for unknown reasons, usually due to a power cut.• Usually indicates a power cut, or other external interruption to the measurement.
• Perform a Clean Optics and Clean Sampling Device protocol before starting a new measurement (see Overview of protocols in manual mode).
E23Temperature Sensor Communication ErrorCommunication with temperature sensor on IO board failed.• Please contact customer service.
E24Laser Communication ErrorCommunication with optical unit failed.• Please contact customer service.
E25Pump Communication ErrorCommunication with pump module impossible.• Please contact customer service.
E26Power Supply Communication ErrorCommunication with power supply module impossible.• Please contact customer service.
E27Pump Communication timeoutCommunication with pump module timed out.• The cartridge door is open.
• A cable is disconnected.
• A circuit board is damaged, either in the pump or cartridge.
E28Empty dye supplyDye tube connected to cartridge is empty.• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E29Empty rinse fluid supplyRinse fluid tube connected to cartridge is empty.• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E30Empty bleach supplyBleach tube connected to cartridge is empty• The tube is disconnected.
• Cartridge is defective.
• Repeat the cartridge change procedure, making sure that all tubes are properly connected.
E31Cartridge electronics failureCartridge level can no longer be determined accurately.• Replace cartridge and contact customer service.
E32Incompatible cartridgeCartridge type is incompatible with this instrument or software.• Make sure that you have the correct cartridge type for this instrument.
E33Processing memory overflowSignal processing circuits memory has overflowed.• Please execute the Clean Optics protocol (see Overview of protocols in manual mode).
• Retry the measurement.
E38Cartridge not ready errorThe new cartridge has not been initialized correctly• Execute the “Cartridge Change” procedure (see Change cartridges).
E39Laser current limit reachedThe laser has reached its upper current limit; indicating end of life.• The laser needs to be replaced. Contact a service technician.
E40Set of gates not foundReference to the set of gates selected for the measurement cannot be found in the instrument settings.• The set of gates has been deleted.
• Change the set of gates and retry the measurement.
E41Set of gates incompatibleThe set of gates selected for the measurement cannot be used with this cartridge.• Correct cartridge change procedure was not followed.
• Incorrect digital input configuration.
• Change the set of gates and retry the measurement.
E43External temperature too lowAmbient temperature is outside of the instrument specifications.• Increase ambient temperature.
• Trying to operate or store the instrument below 5 °C (41 °F) can cause damage to the device.
E47Unable to find main reference magnet during homingDuring initialization, the valve did not home properly.• Please contact customer service.
E53Unable to communicate with motorThe communication to the motor controller of the mixer is defective.• Please contact customer service.
E54Laser temperature sensor errorUnable to get the laser temperature.• Please contact customer service.

16.4 Error and warning logs​

On the Home screen, press Settings, then Logs. Select Error log or Warning log to open the corresponding log.

BactoSense — Error log

Figure 16-1 : Error logs

①

Date and time of the event

②

Type of the event

BactoSense — Critical error log entry

Critical error

BactoSense — Non-critical error log entry

Non-critical error

BactoSense — Log entry: errors cleared

Errors cleared

③

Short description of the error or event

④

Button to manually clear errors (available to Basic, Advanced and Admin users when the instrument is idle)

⑤

Date range of displayed events. Click to change the range.

View event details

Selecting any event on the list leads to a page with more information on the cause of the error and recommended resolution steps.

16.5 Clearing errors​

Critical errors must be cleared manually. Make sure to first solve the origin of the problem, then manually clear the error with the steps below. Please note that the procedure below does not automatically solve the root cause of the error.

WORK STEPADDITIONAL INFO / IMAGE
1.On the Home screen, press Settings to open the settings menu.
2.Press Logs, then Error log.
3.Make sure you have resolved the problem leading to the displayed error(s). Then press the Clear errors button.BactoSense — Clear errors control
4.Cleared errors appear with grey error symbols and an “Errors cleared” event is added to the log. The instrument is ready to analyze again.

16.6 Automatic self-check​

The automatic self-check tool checks that each component of the BactoSense is functioning correctly and identifies those that are not. It can be used to identify problems with the instrument.

WORK STEP

ADDITIONAL INFO / IMAGE

1.

As an Advanced or Admin user, press Settings on the Home screen to open the settings menu.

2.

Press the Maintenance button.

3.

Press the Self-check button.

4.

Additional information

Press the Rerun button. Once execution has terminated, the instrument shows a list of all tests, with a green PASS or red FAIL indicator. Some tests may be skipped: if the system cannot communicate with the cartridge, all the tests that require cartridge communication are skipped.

BactoSense — Self-check results

16.7 Operation at low temperature​

The instrument must be operated within a specific temperature range to prevent damage and ensure reliable and reproducible measurement results (see Technical data). At very low and very high ambient temperatures (below 5 °C (41 °F) or above 30 °C (86 °F)), protocols are not executed. At low ambient temperatures, the instrument automatically warms up prior to a measurement to reach the required temperature. During this heating process, the instrument status shows “Heating...”. If the target temperature cannot be reached after several hours, the protocol exits with an error (table 16-2, E42).

Allow time for heating

Measurements can be delayed if extensive heating is necessary. To avoid this, keep the instrument temperature above 10 °C (50 °F).

16.8 Troubleshooting​

The following table provides an overview of issues and resolutions. If the issue cannot be resolved, please contact your local service representative or bNovate customer service. Contact information for bNovate customer service can be found in Contacting customer service.

Table 16-4 Troubleshooting instructions

OBSERVATIONINSTRUCTION
Nothing on display• Check whether the BactoSense is connected to power and switched on.
Unresponsive or flickering user interface• Potential electromagnetic interference from nearby machine (e.g. pump). Relocate BactoSense.
Error message on the display• Respond to the error message as described in this chapter.
The measurements appear to be wrong• Ensure that the sample and operating conditions described in Technical data are met.
• Check whether the instrument is correctly installed according to Installation.
• Ensure that the servicing duties have been performed according to the servicing schedules described in Periodic preventive maintenance by service professional and Maintenance by the user.
• Perform an instrument cleaning using the Cleaning Kit.
• Perform an instrument validation using the Validation Kit.