3. Safety

3.1 Dangers​

Damaged cables

Damaged instrument or cabling

Touching damaged cables may lead to electrical shocks or death.

  • The instrument may be operated only when the cables are intact.
  • The instrument may be operated only if it has been properly installed or repaired.
Mains voltage

Dangerous voltage inside the supplied mains device connection and cable

  • Touching a wet or damaged mains device may lead to electrical shocks or death.
  • Do not position the mains device in moist locations.
  • Do not operate the mains device if its housing is damaged.
Correct supply voltage

Damage due to incorrect service voltage

If the instrument is connected to an incorrect service voltage, the instrument can be damaged.

  • The instrument must only be connected to voltage sources as specified on the type plate.
  • The instrument must only be operated with the supplied power supply unit, as described in Electrical connections.
Keep the manual available

Missing user manual

Operating the instrument without following the procedures indicated in the user manual may lead to injuries to persons and damage to the instrument.

  • If the instrument changes hands, always include the user manual.
  • If the user manual is lost, you can contact bNovate to request a replacement at www.bnovate.com/contact.
Prevent condensation

Moisture or condensation on the electrical components

If moisture enters the instrument’s electronic enclosure, the instrument can be damaged.

  • The electronic enclosure covers and lids must always be attached during operation.
  • Service inside the electronic enclosure must only be performed by trained personnel in a dry room and at room temperature. The instrument should be at operating or room temperature (avoid condensation on electrical surfaces).
Avoid aggressive chemicals

Use of aggressive chemicals

Use of aggressive chemicals can cause damage to instrument components.

  • Do not use aggressive chemicals or cleaning agents.
  • Should the instrument come in contact with aggressive chemicals, clean it thoroughly with a neutral cleaning agent.
  • Use a damp cloth (water only) to clean the exterior of the device.

3.2 Residual risk​

Measurement accuracy

Risk of the displayed measurement values being incorrect

According to the risk assessment of the applied safety directive DIN EN 61010-1, there remains the risk of the displayed measurement values being incorrect. This risk can be reduced with the following measures:

  • Perform the specified servicing duties.
  • Use an access code on the BactoSense to prevent unauthorized persons from changing parameters.
Cross-contamination

Risk of cross-contamination

According to the risk assessment of the applied safety directive DIN EN 61010-1, there remains a risk of cross-contamination of the water pipes if the check valves inside the BactoSwitch fail. Customers should consider installing mitigation measures (e.g. double check valves) if that risk is not acceptable.

3.3 Warning and danger symbols on the device​

Follow the safety instructions

No warning or danger symbols on the instrument

Users must ensure that they observe the safety measures as specified in the user manual at all times when working with the instrument and its peripheral equipment, even if no warning or danger symbols are attached to the instrument.

  • Observe safety points when performing the described procedures.
  • Observe local safety regulations.