7. Maintenance
7.1 Maintenance by the user
To ensure proper functioning of the device, the following maintenance can be performed by the user.
Avoid contamination of the equipment. Wear nitrile gloves when handling tubings and online sampling devices and any component which comes in contact with them.
Before performing any maintenance operation
- Close the On/Off valves at the water pipe connections.
- Disconnect the mains power supply to avoid any risk of short circuits.
7.1.1 Cleaning
In the event of biological contamination, the BactoSwitch can be cleaned in the following ways:
| AGENT | CONDITIONS |
|---|---|
| Chlorine | Flowing water with up to 10’000 mg/L of active chlorine is allowed. Chlorine cleaning • Wear appropriate personal protective equipment. |
| Hot water | Flowing water at temperatures up to 85 °C, with a reduced maximum pressure of 2 bar, and for up to 15 min. Hot water cleaning • The installation must include a pressure-limiting device at the inlet tubings input, to ensure that the pressure cannot exceed 2 bar. |
Contact support@bnovate.com for questions.
7.1.2 Replacement of input tubings
In the event of persistent biological or non-biological (e.g. iron, manganese) contamination of the tubing, the user can replace it.
Before disconnecting the tubings
- Close the On/Off valves at the water pipe connections.
- Disconnect the mains power supply to avoid any risk of short circuits.
7.2 Preventive maintenance by a service professional
A yearly preventive maintenance intervention must be scheduled to replace wear parts, update software and firmware, and verify proper functioning of the device. This is important to prevent failures, optimize the performance and reliability, and ensure a long product lifetime.
The preventive maintenance must only be conducted by specifically trained service professionals.
Your local service partner can provide you with information about the preventive maintenance schedule and the available options.