7. Field Bus Interface

7.1 Modbus TCP​

7.1.1 General information for operating with Modbus TCP​

The BactoSense can be operated via the Modbus Transmission Control Protocol (TCP). The following requirements must be met:

  • The computer and/or the management or control system must be compatible with Modbus TCP.
  • The computer and/or the management or control system must have software that can properly process the data provided by the BactoSense.
  • The BactoSense must have the Modbus TCP service activated under Home → Settings → System settings → System services.
  • The BactoSense must be connected to the bus system.
  • To be able to work with the Modbus TCP, the network settings must be set correctly.

Additional information

  • The Modbus TCP interface is integrated by default in the BactoSense.
  • The Ethernet socket is located at the right-hand side of the instrument.
  • The address table necessary for programming can be found in the Modbus TCP address list.
  • All registers are encoded big-endian.
  • Modbus TCP communication runs on port 502.
  • The error code lists can be found in Non-critical errors and Critical errors.
  • Modbus TCP can be used for time synchronisation. Follow the instructions in Set NTP Servers and select Modbus TCP as a sync source.

7.1.2 Address list on Modbus TCP:​

The following values can be read with Modbus function 3 or 4 and written with Modbus function 16.

REGISTER
NO
ADDRESS (OFFSET)DATA TYPEFUNCTIONVALUES
300201
400201
0x00C8Unsigned integer bits
0-15
Digital
Inputs.
Ignored unless
“Modbus
TCP” is selected as source in Digital
Inputs settings screen.
Bit 0: Input 1
Bit 1: Input 2
Bit 2: Input 3
Bit 3: Input 4
For correct bit to input mapping, refer to Digital inputs configuration and PLC mode, in particular table 6-2 and figure 6-5.
Input 1111 = 15 means abort.
300202
400202
0x00C9Unsigned integer bits
0-15
Power controlBit 0: Power off
Bit 1: Reboot
300203
400203
0x00CAUint16WatchdogWrite setpoint in seconds to activate or reset countdown.
After the value is written, a countdown decrements it by
1 every second.
During countdown, reading the register returns the setpoint.
If countdown reaches 0, any ongoing protocol is aborted.
Reading the register returns
0
Write 0 to disable countdown.
300204
400204
0x00CBUint16Test registerWrite 1 to display a message on DI source page of GUI.
Resets to 0 at startup.
300211
400211
0x00D2UInt16: YearGet/Set
Date and Time.
Time is changed as soon as write is finished.
Only the fields where registers were written are
changed: if only month is written, only the month is updated.
[0,9999]
300212
400212
0x00D3UInt16:
Month
[1,12]
300213
400213
0x00D4UInt16: Day[0,31]
300214
400214
0x00D5UInt16: Hour[0,24]
300215
400215
0x00D6UInt16:
Minute
[0,59]
300216
400216
0x00D7UInt16:
Second
[0,59]
The following values can be read with Modbus function 3 or 4.
300001
400001
0x0000UInt16 (individual bits)Digital
Outputs
Bit 0: Output 1
Bit 1: Output 2
Bit 2: Output 3
Bit 3: Output 4
For correct bit to output mapping, refer to Digital output configuration, or use the output test interface as described in Digital and analogue output testing.
300002
400002
0x0001UInt16 (individual bits)Instrument
Status
0 = No
1 = Yes
Bit 0: Process running
Bit 1: Critical Error
Bit 2: Non-critical error
Bit 3: PLC mode active
Bit 4: Rapid heating (in winter mode)
Bit 5: Background heating (winter mode)
Bit 6: Sampling device in use
300003
400003
0x0002UInt16 (individual bits)Alarm status, i.e., alarms raised during the last measurement.Bit 0: Alarm 1
Bit 1: Alarm 2
Bit 2: Alarm 3
…
300004
400004
0x0003UInt32 MSB (individual bits)Warning status, i.e., warnings raised during the last measurement.Bit 0: Warning W01
Bit 1: Warning W02
...
Bit 31: Warning W32
300005
400005
0x0004UInt32 LSB (individual bits)
300006
400006
0x0005UInt16Error codeSee error code list. Note:
E00 (unhandled software error) is renamed to 999 to avoid confusion with
"no error". Non-critical error codes remain until either cleared automatically by a successful protocol or cleared manually.
300007
400007
0x0006UInt16HeartbeatSeconds of the clock. The value changes every second, use this to monitor responsiveness of the system.
300008
400008
0x0007UInt16Measurement stageIDLE 0
INITIALIZING 1
HEATING 2
FILLING 3
PRIMING 4
MIXING 5
INCUBATING 6
ANALYZING 7
CLEANING 8
FLUSHING 9
PROCESSING 10
VALIDATING 11
MOVING 12
EJECTING 13
EXPORTING 14
RENAMING 15
DELETING 16
SAVING 17
SELFCHECK 18
ABORTED 19
UNKNOWN 98
ERROR 99
300011
400011
0x000AUInt16Measurement Progress[0, 100]
0 .. 100 %
300012
400012
0x000BFloat 32 MSBTemperature of IO boardCelsius
300013
400013
0x000CFloat 32 LSB
300014
400014
0x000DUInt16Humidity at IO board[0, 100]
0 .. 100 %
300015
400015
0x000EUInt16: YearDate and time of last sampling
300016
400016
0x000FUInt16: Month
300017
400017
0x0010UInt16: Day
300018
400018
0x0011UInt16: Hour
300019
400019
0x0012UInt16: Minute
300020
400020
0x0013UInt16: Second
300021
400021
0x0014UInt16Cartridge percentage[0, 100]
0 .. 100 %
300022
400022
0x0015UInt32 MSBCartridge serial no
300023
400023
0x0016UInt32 LSB
300024
400024
0x0017UInt32 MSBFill serial no
300025
400025
0x0018UInt32 LSB
300026
400026
0x0019Uint16Cartridge type1 = TCC
1002 = ICC
1030 = ACC
1003 = ICC-BWT
300027
400027
0x001AUInt16: YearCartridge expiration dateUTC date
300028
400028
0x001BUInt16: Month
300029
400029
0x001cUInt16: Day
300031
400031
0x001EUInt32 MSBInstrument
Serial number
300032
400032
0x001FUInt32 LSB
300033
400033
0x0020String char 1-2Software versionWhen shorter than 16 characters, the string is padded with null
characters. Example:
'1.4.0-BS\0\0\0\0...'
300034
400034
0x0021String char 3-4
300035
400035
0x0022String char 5-6
300036
400036
0x0023String char 7-8
300037
400037
0x0024String char 9-10
300038
400038
0x0025String char 11-
12
300039
400039
0x0026String char 13-
14
300040
400040
0x0027String char 15-
16
300041
400041
0x0028UInt16Memory remainingApprox measurements remaining before disk is full
300042
400042
0x0029UInt16Lifetime of valve remaining[0, 100], percentage of movements remaining before service.
300043
400043
0x002AUInt16Lifetime of plunger remaining[0, 100], percentage of movements remaining before service.
300044
400044
0x002BUInt16: YearNext service due dateUTC date
300045
400045
0x002CUInt16: Month
300046
400046
0x002DUInt16: Day
300101
400101
0x0064UInt32 MSBTCCResets to zero after a cleaning protocol or at reboot.
300102
400102
0x0065UInt32 LSB
300103
400103
0x0066UInt32 MSBICCAlways 0 if cartridge does not provide ICC.
300104
400104
0x0067UInt32 LSB
300105
400105
0x0068UInt32 MSBGate+Always 0 if cartridge does not provide Gate+
300106
400106
0x0069UInt32 LSB
300107
400107
0x006AUInt32 MSBHNAC
300108
400108
0x006BUInt32 LSB
300109
400109
0x006CUInt32 MSBLNAC
300110
400110
0x006DUInt32 LSB
300111
400111
0x006EFloat 32 MSBHNAP
300112
400112
0x006FFloat 32 LSB
300113
400113
0x0070UInt32 MSBACCAlways 0 if not ACC cartridge.
300114
400114
0x0071UInt32 LSB
300115
400115
0x0072UInt32 MSBHACCAlways 0 if not ACC cartridge.
300116
400116
0x0073UInt32 LSB
300117
400117
0x0074UInt32 MSBLACCAlways 0 if not ACC cartridge.
300118
400118
0x0075UInt32 LSB
300119
400119
0x0076Float 32 MSBHACPAlways 0 if not ACC cartridge.
300120
400120
0x0077Float 32 LSB
300121
400121
0x0078UInt32 MSBLive TCCSame as TCC, but the value is updated live during a measurement.
Useful for early contamination detection
(for example: halting a pump or diverging online water).
Returns to zero at the end of the measurement.
300122
400122
0x0079UInt32 LSB
300123
400123
0x007AUInt32 MSBLive ICCsee above.
300124
400124
0x007BUInt32 LSB
300125
400125
0x007CUInt32 MSBLive Gate+see above.
300126
400126
0x007DUInt32 LSB
300127
400127
0x007EUInt32 MSBLive HNACsee above.
300128
400128
0x007FUInt32 LSB
300129
400129
0x0080UInt32 MSBLive LNACsee above.
300130
400130
0x0081UInt32 LSB
300131
400131
0x0082Float 32 MSBLive HNAPsee above.
300132
400132
0x0083Float 32 LSB
300135
400135
0x0086UInt32 MSBLive ACCsee above.
300136
400136
0x0087UInt32 LSB
300137
400137
0x0088UInt32 MSBLive HACCsee above.
300138
400138
0x0089UInt32 LSB
300139
400139
0x008AUInt32 MSBLive LACCsee above.
300140
400140
0x008BUInt32 LSB
300141
400141
0x008CFloat 32 MSBLive HACPsee above.
300142
400142
0x008DFloat 32 LSB
300151
400151
0x0096UInt16: YearDate and time of last abrupt shutdown.
300152
400152
0x0097UInt16: Month
300153
400153
0x0098UInt16: Day
300154
400154
0x0099UInt16: Hour
300155
400155
0x009AUInt16: Minute
300156
400156
0x009BUInt16: Second
300157
400157
0x009CUInt16Power off timeIn seconds
300165
400165
0x00A4UInt16: YearDate and time of last protocol start.
300166
400166
0x00A5UInt16: Month
300167
400167
0x00A6UInt16: Day
300168
400168
0x00A7UInt16: Hour
300169
400169
0x00A8UInt16: Minute
300170
400170
0x00A9UInt16: Second
300171
400171
0x00AAUInt16BactoSwitch current measurement streamFrom 1 to 4
300172
400172
0x00ABUInt16BactoSwitch next measurement streamFrom 1 to 4
300173
400173
0x00ACUInt16BactoSwitch flow ratemL/min
300174
400174
0x00ADUInt16BactoSwitch status0 means no error.
300175
400175
0x00AEUInt16BactoSwitch connection0 = not connected
1 = connected successfully
300176
400176
0x00AFUInt16BactoSwitch valve stateBitmask 0x01 = valve of stream 1, etc int 15 = all open
300181
400181
0x00B4UInt32 MSBCBC
300182
400182
0x00B5UInt32 LSB
300183
400183
0x00B6UInt32 MSBEBC
300184
400184
0x00B7UInt32 LSB
300185
400185
0x00B8UInt32 MSBNBC
300186
400186
0x00B9UInt32 LSB
300187
400187
0x00BAUInt32 MSBLive CBCsee above.
300188
400188
0x00BBUInt32 LSB
300189
400189
0x00BCUInt32 MSBLive EBCsee above.
300190
400190
0x00BDUInt32 LSB
300191
400191
0x00BEUInt32 MSBLive NBCsee above.
300192
400192
0x00BFUInt32 LSB